Biphasic Defibrillator

Technical Specifications

General

Display: 7″ Colour TFT
Display area: 150mm x 90mm
Dimensions: H280 x W 330 x D220 mm (Max) without Paddle
Weight: Approx 6 Kgs.
Unit ON-OFF: By front panel keyboard, Encoder
Operation: User adjustable high & low limit, Audible and Visual alarming
Alarm: 24hrs Graphical and Tabular trend of ECG, SpO2, NIBP EtCO2
Charge Discharge Control: Charge and Shock button on panel and paddles.

Defib (Internet Paddles) (Optional)

Operating Mode: Manual
Waveform: Biphasic Truncated Exponential
Energy: 2 to 50 Joules in 8 steps selection
Synchronous Cardie Version: Energy delivery begin with in 60 ms of the QRS Peak
Charging time: 6 Seconds on Mains
Paddle Assembly: Internal paddle assembly with paddle extension cord

Printer

Paper Type: Thermal paper 50/55mm width, 20 meters length for 2nd printer
Event Printing: Prints 5 Seconds pre & post ECG with Date, Time, Energy and Mode, Shock status
ECG in Defib Mode: Prints last 10 sec ECG through paddle
Prints Annotates in defib Mode: Time, Date, Mode, lead name, Guin, Sweep speed
ECG in Monitor Mode: Prints last 10 sec ECG through cable Prints Annotates in Monitor Time, Date, selected Lead name, Guin, Sweep Speed
Manual Defib Check Mode: Prints the status of shock, Energy, Date & Time

AED Mode

Waveform: Biphsic truncated Exponential
AED Shock Series: Energy level 200J
Charge command: Automatic after identifying shockable rhythm
Shock command: Front panel button, Shock
Controls: Analyze
Defibrillation Electrode: Self Adhesive, adult pads
Voice and Text prompts: Extensive text & audible messages to guide user

Environmental

Operating Temp: -5 C to +50 C
Humidity: Up to 95% non-condensing

Power Supply

Input: 150-276 VAC, 50Hz, 1 O
Battery: 12 V 4.5Ah, Rechargeable, (Sealed Lead Acid)
Back-up Time: 3 hours min, on full charge in ECG mode 70 – 100 shoc of 150 joules delive (Full Charge)
Power Consumption: 30 Watts(max.) in ECG mode

Heart Rate

Range: 10 – 250 BPM
Accuracy: +/-2 BPM
Resolution: 1 BPM
Tone: Beep on every QRS detected
Alarm Limit setting: 10-250 (step of 1)

Defib (External Paddles)

Operating Mode: Manual
Waveform: Biphasic Truncated Exponential
Energy: 20 to 200 joules in 12 steps selection
Energy display: Display selected and charged/delivered energy
Synchronous Cardio Version: Energy delivery begins with in 60 ms of the QRS Peak
Charging Time: 6 seconds on Mains
Paddle Assembly: External paddle assembly with coil retractable high voltage cable, Adults & Pediatric paddle integrated to same handle
Check Mode: Manual check mode
Patient Impendence range: 35 ohm to 250 ohm
Sock Delivery: Through paddles or pads

ECG

Frequency response: 0.5 Hz to 35Hz
Patient isolation: Optical isolation
Lead Selection: Lead I, II, and III
Gain setting: 0.5, 1.0 and 2.0
CMRR: >90 dB
Sweep Speed: 12.5, 25.0, 50.0 mm/sec
Digital Filter: Digital signal processing for good quality of ECG
ECG traces Viewing: 4 seconds ECG traces can be viewed