HRP Fully Automatic Analyzer
General Specifications
Item Description
Throughput 100 test/hour with 1/2 reagent
System type Discreet. open, fully automated, patient prioritized, selectable 1/2 reagent system
Sample Serum

Urine

Plasma

CSF

Others

Measurement principle Latex Enhanced Turbidimetric Immunoassay

Turbidimetric Immunoassay

Photometric (Rate/end point)

Onboard Reagent Positions 25 positions of 20 ml bottles for RI/R2
Programmable parameters 50 photometric tests
Assay modes End Point

Fixed time (2-point)

Kinetic Rate-A

Kinetic Rate-B

Sample volume 2-50 μl (adjustable in 0.1 μl step)
Reaction volume 180μl to 5ooμl
Reaction temperature 37 •c ± 0.2 •c
Reaction time Single reagent chemistry.30-600sec,

Double reagent chemistry: 30-300sec

Calibration Maximum 20 calibrators can be used

6 calibrators maximum per test

On-board auto dilution Of Calibrator

Mixing type Variable speeds paddle mixer (Low, Middle, High)
Maintenance Auto scheduling — Start of Day, End of Day
Identification Position Identification for samples and reagents

Fixed Bar code reader for Identification for samples

Water supply unit NCCLS Type 2: Up to 4.5 liters/hour (approx)
Quality control Up to 3 levels programmable for every parameter
System Warm-Up 10 minutes system warm-up time.
Level sensing and mechanism Vertical & horizontal obstruction detection Capacitance based liquid level sensing
Noise level Less than 65 dB with cover closed
CPU & Display P-IV with ( IGB)RAM, 40 GB or Higher Hard Drive,

CD-RW suitable WIN OS

15 inch TFT color display

Keyboard Standard keyboard & standard mouse
Printer External: A4 size Desk Jet color printer (Optional)
System interface Analyzer — PC: RS-232C bi-directional, PC — Printer. USB (Min 2 Ports) & a Serial Port

PC — Host computer via serial RS232 or TCP/IP

Sample/Reagent unit
Item Description
Sample tray Rotating Turn Table tray
Sample positions 25 positions
Reagent positions 25 positions
Sample container Blood collection tube Of 5 ml volume
Standard cup 2 ml
Reagent bottles 20ml bottles
Reagent cooling 15 deg C below ambient room temperature (approx 25 deg C)
STAT samples Can be placed at any free position
Reagent volume Reagent 1: 180 – 350 μl (adjustable in 1 μl step) Reagent 2: 10 – 350 μl (adjustable in 1 μl step)
Sample/Reagent Probe Micro-pipette with level sensor
Sample dilution Dilution ratio: 2 to 40 times
Real time reagent inventory Displayed on screen in ml and Number of Tests
Turntable protection Cover protection from evaporation, dust. and direct light
Carry over evasion Wash with System water/ Reagent/Wash Solution
Cuvette washing Number Of washing operation steps: 7 steps

Reaction waste removal: 1 step in CRU.
DI water. 4 steps
Residual liquid removal: 2 steps

Optical unit
Item Description
Photometric system Multiple wavelength, Maintenance free sealed filters with optical fibers
Wavelength range 9 Wavelenghts:
340, 405, 450, 510, 546, 578, 620, 660, 690
Wavelength selection Mono or Dual (User selectable)
Cycle time 30 seconds
Path Length 5 mm
Resolution 0.0001 OD
Photometer Linearity Up to ± 3.0 Optical Density
Light source Tungsten halogen lamp (12V 20 W)
Detector Silicon photo-diode
Cuvette blanking Water blank after the cuvette washing
Control Unit
Item Description
User interface hardware PC: Windows machine, IBM compatible

OS: Windows XP (Service Pack 2)/Windows 7

CPU: Pentium IV, RAM: ( IGB), Hard disk: 40GB

Console: 15 inch TFT color monitor

Multimedia: Speakers with sound card

External drives: CD-ROM drive

System interface Analyzer – pc: RS-232C bi-directional (RS232-USB Converter)
Printer Option External: A4 size color printer
Data Processing
Description
Calibration curves K-Factor, Linear (one. two point and multipoint),

Logit-log (4P, 5P), Spline, Exponential, Polynomial

Calibration history — last 10 for each parameter Flexibility of using previous factors

Quality control Daily and Monthly L-J Graph plots

Mean, SD, %CV, R is calculated for each chemistry

Real time quality control based on Multi-QC-rules

Sample Scheduling Possible to make profiles for speedier scheduling

Possible for sample replication for same test

Copy scheduling of 1 sample to other samples

Auto Re-run Auto Rerun with sample dilution for greater than Linearity samples
Monitor function Offline/Online reaction curve display
Calibration curve graphical display
Operation status by run monitor
Correlation Equation Correlation correction factor (Y = ax + b)
Data storage capacity Test results: 1,000,000
Reaction curve: 40,000 tests
Report/list format Unique Alphanumeric patient id
Report search: Patient wise, Test wise, Date wise, Doctor/Hospital wise etc.
Blanking correction Sample Blank and Reagent Blank correction
Data check Reference range check by age, sex. sample type
Normal and high reaction linearity
Reaction mixture absorbance checks
Password Dual level access passwords
Installation Conditions
Item Description
Power source/ consumption AC 220 V ± 10%, 50 Hz or AC 110 V ± 10%, 60 Hz Power consumption: 2000 VA
Water supply De-ionized water consumption: 4.5 liter/hour
Ambient temperature Approx 25 •C
Relative humidity 40 — 80% free from water dew formation
Dimensions 550 (W) x 420 (D) x 380 (H)
Weight Approximately 30 kg.
Accessories/Consumables/Spares
Item Description
Installation parts Cuvette set
Operator Manual
Application Software CD
Reagent tray Reagent bottles
Waste reservoir
Waste connection tubes
Connection cables kit
Consumables Sample cup/tubes
Spares Cuvettes
Cuvette Drier
Photometer Lamp Assembly
Set of Fuses